Output Adjustment (Out) - Presys MCS-12-IS Technical Manual

Universal process calibrator
Table of Contents

Advertisement

3.2. Output adjustment (OUT)

52
The output adjustment (except for CJC and Probe) is performed in STEPS. For each STEP the calibrator
outputs a signal of the same type selected which must be measured and stored.
It is also possible to make a simple verification in the last two STEPS of each selected range, further details
will be given below.
1) V, mV e mA.
For these output ranges, the display shows three information fields.
Field (1) is the current step. To go to the next step, press  and to go backwards, press . In each step, the
calibrator outputs the signal automatically.
Field (2) is the intimal value measured by the calibrator, expressed in percentage (%) of the range. Before
storing the adjustment values, one should wast a few seconds for output signal stabilization.
Field (3) is the value intered by the user. After output signal measurement, press "ENTER" followed by the
value measured and "ENTER" again.
The adjustment of these ranges comprises 5 steps. In steps STEP1 to STEP3, the signals are output and
should be measured and stored, by pressing "ENTER". In steps STEP4 and STEP5, signals are output only for
verification purpose and no values are stored. The verification points are listed in the table below.
STEP 1 (1)
11.82813 (3)
88.7% (2)