2.9.1. SCALE Function

40
It establishes a linear relationship between the calibrator input signal and what is shown at the display,
according to the graphic below.
The scaled indication at the display (#) may represent any engineering unit, such as: m/s, m
The number of decimals, up to 4, shown at the display may be configured by using parameter Scale Dec.
The value for Input High must be necessarily higher than Input Low. On the other hand, Scale High and
Scale Low may have any relationship between themselves: higher than, lower than or equal to, and they may have a
signal before them. Thus direct or reverse relationships may be established.
The counter and the contact inputs may not be scaled.
Fig. 10 - SCALE Function (LINEAR).
³
/s, %, etc.