Adjusting procedures for this calibrator must be performed under the reference conditions of temperature and
humidity.
Better calibration results are achieved if warm-up time is at least two hours and if the battery charger remains
disconnected from the calibrator since one hour before its usage.
The standards used to adjust this calibrator should have accuracy at least 3 times better than the accuracy values
provided in this manual.