Adjusting procedures for this calibrator must be performed under the reference conditions of temperature and
•
humidity.
•
Better calibration results are achieved if warm-up time is at least two hours and if the battery charger remains
disconnected from the calibrator since one hour before its usage.
•
The standards used to adjust this calibrator should have accuracy at least 3 times better than the accuracy values
provided in this manual.