The PROBE option is used to configure the parameters using the Callendar-Van Dusen curve. Entering this
menu, the options will appear.
The IEC751 option contains the standard table for conversion between resistance and temperature,
according to the IEC-60751 standard. When not using a custom correction curve, leave this option selected.
The CUSTOM option allows the configuration of CVD parameters. These parameters belong to the equation
described below.
R(t) = R0.{1 + A.t +B.t
Where t refers to temperature on this scale and R0 to resistance at 0 ºC.
Set the value of R0 and the coefficients A, B and C, together with their respective exponents in E_A, E_B
and E_C.
IEC751
2
+ C.t
3
(t-100)}, C = 0 for t ≥ 0 ºC.
 CUSTOM