Probe
12
Independent input for RTD (Probe). Probe is a high accuracy 4-wire Pt100 available under previous consult.
For a higher level of accuracy, it is possible to configure the parameters of the sensor Callendar-Van Dusen
curve in the MCS-12-IS, correcting the errors found in the sensor certificate (see section 3.8.3 - Probe).
Software Special Functions
- Any output programmable in:
1) STEP: 10%, 20%, 25% or up to 11 programmable setpoints via key or adjustable time.
2) RAMP: up and down with programmable travel and dwell time.
- Special functions:
1) SCALE: makes the scaling of both input and output in 6 digits and allows the configuration of decimal point.
2) CAL: scales the input in same unit of output.
3) CONV: converts any input into any output, galvanically isolated.
- MEM command: It can store up to eight types of configuration chosen by user.