To remove an O-ring:
Notice:
ensure that you do not scratch the component with the removal tool.
Use the tools as aids to pull the O-ring or seal from its groove.
Tip:
If you do not plan to reuse the O-ring or seal, you can use the forked end of tool 1 to impale
the O-ring or seal to remove it.
Warning:
To avoid spreading contamination with biologically hazardous, toxic, and corrosive
materials, dispose of all waste materials according to local environmental regulations.

4.8 Cleaning the instrument case

Notice:
solvents when cleaning them.
Use a soft cloth, dampened with water, to clean the outside surfaces of the mass spectrometer.

4.9 Emptying the nitrogen exhaust trap bottle

Inspect the nitrogen exhaust trap bottle in the instrument exhaust line daily and empty it before it
is more than approximately 10% full.
To avoid damaging the component when removing an O-ring or seal from it,
To avoid abrading the surfaces of the instrument, do not use abrasives or
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)
Page 83