Removing The Differential Aperture Support And The Differential Aperture From The Ion Guide Assembly - Waters Xevo TQ-S cronos Overview And Maintenance Manual

Table of Contents

Advertisement

2. Grasping the ion guide assembly by its top and bottom circuit boards, carefully remove the
differential aperture and ion guide assembly from the pumping block assembly.
4.14.3 Removing the differential aperture support and the differential
aperture from the ion guide assembly
Required materials
• Chemical-resistant, powder-free gloves
• 2.5-mm hex wrench
• Flat-blade screwdriver
• O-ring removal kit
Warning:
To avoid personal contamination with biologically hazardous, toxic, and corrosive
materials, wear chemical-resistant, powder-free gloves when performing this procedure.
1. Place the ion guide assembly with differential aperture onto a clean, flat surface.
Figure 4–26: Removing the differential aperture support and differential aperture
Ion guide assembly
Differential aperture support and circuit board
Securing screws (2 per side)
Differential aperture
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)
Page 109