4 Maintenance Procedures

Cleaning the instrument case

Notice:
clean the instrument's case.
Use a soft cloth, dampened with water, to clean the outside surfaces of the instrument.

Emptying the exhaust trap bottle

Inspect the exhaust trap bottle in the instrument's exhaust line daily, and empty it before it
is more than 10% full.
Nitrogen exhaust trap bottle:
From instrument
waste (12-mm)
108
January 11, 2016, 715004599 Rev. B
To avoid damaging the instrument, do not use abrasives or solvents to
To laboratory
exhaust (12-mm)
Valve control
cable (from
instrument)