Figure 4–4: O-ring removal kit:
Tool 1
Tool 2
To remove an O-ring:
Note:
To avoid damaging the component when removing an O-ring or seal from it, ensure that
you do not scratch the component with the removal tool.
Use the tools as aids to pull the O-ring or seal from its groove.
Tip:
If the O-ring or seal will not be reused, you can use the forked end of tool 2 to impale the O-
ring or seal, aiding its removal.

4.9 Cleaning the instrument case

Notice:
solvents when cleaning them.
Use a soft cloth, dampened with water, to clean the outside surfaces of the mass spectrometer.

4.10 Emptying the nitrogen exhaust trap bottle

Inspect the nitrogen exhaust trap bottle in the instrument exhaust line daily, and empty it before it
is more than approximately 10% full.
To avoid abrading the surfaces of the instrument, do not use abrasives or
June 9, 2016, 715004990 Rev. A
Page 101