4. Install the impactor pin in the UniSpray source (see
impactor pin (Page

4.25 Replacing the instrument's fuses

Warning:
supply before replacing fuses. The mass spectrometer has two fuses and uses double
pole/neutral fusing circuitry. Circuits can remain live even when one fuse has blown.
Warning:
ratings affixed to the rear panel of the module.
If one or both of the mass spectrometer's fuses blow, the instrument will shut down immediately.
If this occurs, disconnect the power cord from the rear panel, and replace the fuses, located at
the bottom left-hand-side of the instrument's rear panel, with T10AH250V, 6x32mm fuses.
138)).
To avoid electric shock, disconnect the mass spectrometer from the power
To avoid electrical fire, ensure that replacement fuses comply with the
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)
Removing and installing the UniSpray
Page 140