Replacing The Esi Probe Tip And Gasket; Removing The Esi Probe Tip And Gasket - Waters Xevo TQ-S cronos Overview And Maintenance Manual

Table of Contents

Advertisement

Figure 4–36: PEEK fitting
PEEK fitting
3. Remove the probe assembly.
Warning:
corrosive materials, dispose of all waste materials according to local environmental
regulations.
4. Dispose of the probe assembly in accordance with local environmental regulations.
5. To install a new probe assembly, see

4.17 Replacing the ESI probe tip and gasket

Replace the ESI probe tip if a blockage occurs in the internal metal sheathing through which the
stainless steel capillary passes or if the probe tip is damaged.

4.17.1 Removing the ESI probe tip and gasket

Required materials
• Chemical-resistant, powder-free gloves
• 7-mm open-end wrench
• 10-mm open-end wrench
To avoid spreading contamination with biologically hazardous, toxic, and
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)
Installing the probe assembly (Page
Page 122
54).