5 How to measure with the measuring module
5. Connect the measuring module to the current path of the load whereby the
blue connection cable must be connected to the positive pole and the blue
socket (A). The black connection cable must be connected to the ground
(COM).
6. Switch on the load.
➢ The measured values will be displayed on the ALMEMO
5.3 How to measure voltage AC with the module ZAD
903-ABx
Wire in a zero-potential condition (especially for measurements above
50V). Switch on the voltage afterwards.
During operation never touch any exposed connections and
components to avoid electric shocks.
Use the supplied touch protected connection cables.
1. Switch off the voltage on the voltage source, and check that no voltage is
applied to the module.
2. Insert the ALMEMO
socket of the ALMEMO
3. Switch on the ALMEMO
- 16 –
plug of the measuring module into an ALMEMO
®
measuring device.
®
measuring device.
®
measuring device.
®