8.2
Access service menu
Parameters which can be changed in Service Mode are protected through verification. On a verified crane
scale, the verification seal must first be destroyed in order to activate Service Mode. It is then necessary
to re-verify the crane scale.
Entering the service menu, small construction
Entering the service menu, large construction
Operation in service menu
Navigation in the service menu is the same as in the user menu, see chapter 7.3.
* To do this on verified scales the lead seal must be destroyed; the verification status is
then no longer valid, see chapter 8.4.
HTS-BA-e-1024
1. Remove the sealing screw (1) *.
2. Switch on the scale.
3. Press the switch in the hole. The display
shows F1.
The scale is ready for modifying parameters.
1. Remove the sealing screw (1) *.
2. Switch on the scale.
3. Press the switch in the hole. The display
shows F1.
The scale is ready for modifying parameters.
51