Battery Depleted During Use, Limited Mode (Cpr); Pads Expired - Schiller FRED easyport plus User Manual

Automated external defibrillator (aed)
Table of Contents

Advertisement

®
®
FRED
easyport
plus
Battery Low
Shock disabled low bat
Fig. 3.2
Depleted battery indication
Configuration
Selftest
>>>
Start
Bluetooth pairing
Device information
Device Settings
Device Update

Pads Expired

>>>
Production
>>>
>>>
Log Files
User Guide
3.2.3

Battery depleted during use, limited mode (CPR)

 Patient hazard — Defibrillation is no longer possible if a depleted battery is
detected. The battery needs to be replaced immediately.
If a depleted battery is detected while the device is in use, the device will show after
the "Low battery" notification the "Shock disabled low bat" notification. An audible
signal is emitted and the battery symbol on the display is blinking red.
 Immediately shut down the device, disconnect electrode connector and replace
the battery.
3.2.4
Pads expired
To monitor the expiry date of the defibrillation pads, always enter the expiry date
printed on the packaging. A notification "Pads expired" appears when the selftest
detects that the entered data has expired. Replace the electrodes as soon as
possible.
Procedure
1.
Press and hold the button
2.
Go to menu "Pads expired" select Year and month and enter the expiry date from
the defibrillation pads packaging.
Prev
– Verify the entered date in the menu
>>>
Pads
>>>
OK
>>>
Next
Expired.
Initial operation
Battery monitoring
while switching the device on
Configuration > Device information >