Battery Depleted During Use, Limited Mode (Cpr) - Schiller FRED easy User Manual

Automated external defibrillator (aed)
Hide thumbs Also See for FRED easy:
Table of Contents

Advertisement

3
Initial operation
3.2
Battery monitoring
Green indicator is off
Page 26
3.2.3

Battery depleted during use, limited mode (CPR)

Patient hazard — Defibrillation is no longer possible if a depleted battery is detected.
The battery needs to be replaced immediately.
If a depleted battery is detected while the device is in use, a message is displayed
prompting the user to replace the battery and to perform CPR. An audible signal is
emitted and the indicator remains off until the battery is replaced (or charged).
The message stays displayed until the battery is replaced.
Depleted battery during self-test
• When a depleted battery has been detected, the green indicator is off and an au-
dible signal is emitted.
• When the device is next switched on, a message is displayed, prompting the user
to replace the battery and continue with CPR until the device is fully operational
again.
Depleted battery after the battery has been inserted
• If a battery inserted in the device is identified as depleted, an audible signal is emit-
ted, the green indicator is turned off and a message indicating that the battery must
be replaced is displayed.
REPLACE BATTERY
30 CHEST COMPRESSIONS THEN
2 RESCUE BREATHS
FRED easy