Alan Bentley
June 19, 2025
With the handle facing away from me the bottom red is LED flashing. The pads are due to expire in July but i have put new pads now the LED red light is still flashing. The machine is 3 years old the battery expiry date is 2031.
Components and Operation ....12 General Information ............12 Design.................. 13 2.2.1 Overview of the configurable settings ..........14 Operating and Display Elements ........15 Overview DefiSign Life AED .............15 2.3.1 2.3.2 Display and Operating Elements............16 Function ................17 Initial operation ........18 Inserting the battery ............
Page 4
DefiSign Life AED Defibrillation ..........24 Instructions and Safety Notes ........... 24 4.1.1 Instructions ..................24 4.1.2 Safety notes for AED use..............24 Applying the adhesive electrodes........26 4.2.1 General information ................. 26 4.2.2 Unpacking and applying electrodes ..........26 4.2.3 Applying the electrodes to the patient’s chest........
Page 5
DefiSign Life AED User Guide Technical Data ........47 System Specifications ............47 Classification and safety standards ......... 48 Defibrillation pulse ............. 49 7.3.1 Shock Advisory System ..............51 Electromagnetic interferences .......... 52 7.4.1 Electromagnetic emissions .............. 52 7.4.2 Electromagnetic immunity ..............52 7.4.3...
Responsibility of the User ® Regulations on who is allowed to use devices like the DefiSign Life AED and which training is required, are country-specific. In any case, legal regulations have to be observed.
25 %. The DefiSign Life AED must not be used in ambulances and emergency vehicles. Organisational Measures Before using the unit, ensure that an introduction regarding the unit functions and the safety precautions has been provided and understood.
Do not, under any circumstances, immerse the device or cable assemblies in liquid. To ensure patient safety, only use original SCHILLER accessories. The user is responsible for the use of third-party accessories. The warranty does not cover damage resulting from the use of accessories or consumables other than those marketed by SCHILLER.
There are no express or implied warranties which extend beyond the warranties here- inabove set forth. SCHILLER makes no warranty of merchantability or fitness for a particular purpose with respect to the product or parts thereof.
BF symbol. The device's signal input is defibrillation-protected. Caution! High voltage! CE-0459 marking (notified body LNE/G-MED). ® Do not dispose of the DefiSign Life AED and its accessories in the household waste. Manufacturer symbol, manufacturing date. Observe the user guide...
Safety Notes DefiSign Life AED 1.11 Symbols/Indicators 1.11.3 Symbols used on the battery The battery is recyclable Do not recharge Do not short-circuit Do not incinerate Do not cut Do not crush Normal storage temperature duration and allowed out-of-range temperature dura-...
Safety Notes DefiSign Life AED User Guide Symbols/Indicators 1.11 1.11.4 Symbols used on the electrode packaging • Remove the patient's clothes • Open the electrode packaging • Peel off the protective foil Disposable item; do not reuse Do not bend packing...
AEDs to an Emergency Medical Technician or First Re- sponders after they have undergone special training. Highly frequented areas are typical places for the operation of a DefiSign Life ® . Some examples below: –...
The device can be provided with different languages. Optional configuration with 3 languages, selectable after switching the device on. ® Metronome The DefiSign Life AED sets a configurable pace for the cardiopulmonary resuscitation (CPR). FreeCPR (option) Information on the chest compression frequency using impedance variation acquired with defibrillation pads.
These modifications need to be registered in the device documentation as well as communicated to all users. SCHILLER's service centre can configure the following parameters: Configurable parameters • Selection of the default language at device start •...
Components and Operation DefiSign Life AED User Guide Operating and Display Elements Operating and Display Elements 2.3.1 Overview DefiSign Life AED Service status LEDs: • Modem • Battery Main status LED • Service • Electrodes Battery The device is switched on when the cover is opened.
Components and Operation DefiSign Life Operating and Display Elements 2.3.2 Display and Operating Elements In addition to the voice prompts, the resuscitation steps are indicated by pictograms and the current step is highlighted with a flashing LED. Basic device with one language As soon the cover of the device is opened, the device starts issuing acoustic advices.
Function Function ® Immediately after a battery has been inserted, the DefiSign Life AED performs a test of the device and battery. If this test is completed successfully, the green status LED is blink-ing and all service status LEDs are off, showing that the device has not detected an error.
Inserting the battery 3 Initial operation ® Danger of explosion — The DefiSign Life AED must not be used in areas where there is an y danger of explosion. Areas may be susceptible to explosion if flammable substances (gas), flammable anaesthetics, or products used to clean or disinfect the skin are used.
Insert the two stop blocks located at the bottom of the battery in the device slots. Perform a rotational movement until the battery locks in place. ® As soon as the battery is inserted, the DefiSign Life AED runs a self-test to check the condition of the device and the battery.
Initial operation DefiSign Life AED Inserting the battery 3.1.2 Switching device On and Off Switching ON Open the cover. The 3 LEDs for the resuscitation steps are briefly lit. Switching OFF Close the cover Forced shutdown procedure If the device cannot be switched off via the above procedure, remove the battery and insert it again.
The old battery must be recycled in accordance with local regulations. 3.2.1 Sufficient battery capacity ® The main status LED (green) on the DefiSign Life AED is blinking when the battery capacity is sufficient to perform the resuscitation protocol. 3.2.2 Low battery capacity indication •...
Initial operation DefiSign Life AED Battery monitoring 3.2.3 Battery depleted during use, limited mode (CPR) Patient hazard — Defibrillation is no longer possible if a depleted battery is detected. The battery needs to be replaced immediately. If a depleted battery is detected while the device is in use, the device will prompt the user to replace the battery and perform CPR.
User Guide Replacing the “pre-connected” pads Replacing the “pre-connected” pads ® The DefiSign Life AED is delivered with "pre-connected" pads. To replaced the pads after use or if the shelf life has expired, proceed according to the following instructions: Only use the pads up to their expiration date.
Improper use can endanger life. • Non medical personnel is only permitted to use an AED such as the DefiSign Life ® if local law approves of this practice.
Page 27
Defibrillation DefiSign Life AED User Guide Instructions and Safety Notes Shock hazard — for user and assistants Position the patient flat on a firm, electrically insulated surface. Make sure that there are no conductive connections between the patient and other persons during ECG analysis and defibrillation.
The pre-connected electrodes are stored in the defibrillator cover and can be ac- cessed when the cover is opened. A spare set of adult or children electrodes can be found in the compartment on the bottom of the DefiSign Life AED. 4.2.2 Unpacking and applying electrodes ...
Defibrillation DefiSign Life AED User Guide Applying the adhesive electrodes 4.2.3 Applying the electrodes to the patient’s chest Skin covered in sea water, sand, sunscreen, or skin or body care products may impair electrode contact or cause the electrodes to become disconnected.
Page 30
Defibrillation DefiSign Life AED Applying the adhesive electrodes Children weighing less than 25 kg (younger than 8 years of age) Electrode placement on children weighing less than 25 kg (see Fig. 4.5 Application sites for children weighing less than 25 kg).
Defibrillation DefiSign Life AED User Guide Applying the adhesive electrodes 4.2.4 Checking the electrodes If the resistance (impedance) reaches an unacceptable value, the device interrupts and prompts the user to check the electrode application; in addition, the green indica- tor is blinking This can occur if: –...
Defibrillation DefiSign Life AED Semi-automatic defibrillation Semi-automatic defibrillation Patient hazard — The guidelines given in 4.1 Instructions and Safety Notes must be observed. Semi-Automatic Defibrillation Depending on the configuration of the device, the instructions provided by the device may be shortened.
Page 33
Defibrillation DefiSign Life AED User Guide Semi-automatic defibrillation Step 3 Shock delivery When the energy is charged, the user is prompted to trigger the shock by pressing the blinking orange button. Shock hazard! Do not, under any circumstances, touch the patient during shock delivery.
Defibrillation DefiSign Life AED Automatic defibrillation Automatic defibrillation The laws and regulations for the use of automatic defibrillators vary from country to country. While some countries allow laypersons to use automatic defibrillators without any special training, other countries restrict the use of AEDs to EMTs or First Re- sponders who have undergone special training.
Page 35
Defibrillation DefiSign Life AED User Guide Automatic defibrillation Automatic defibrillation Step 1 Switching on and preparing the device Open the cover to switch the device on. – If the cover is missing, remove the battery and insert it again to switch the device Assess the patient's condition: not responsive, no respiration, no pulse.
Page 36
Defibrillation DefiSign Life AED Automatic defibrillation Step 3 Automatic shock delivery As soon as the energy charging is completed, the device automatically delivers the shock, without user intervention. An acoustic countdown starts and the orange button blinks until the shock is delivered.
Defibrillation DefiSign Life AED User Guide Internal safety discharge Internal safety discharge If the device's behaviour differs from the description given in this user guide, the device is defective and must be repaired. An internal safety discharge ensures that the stored energy is discharged within the device every time a defibrillation shock was not delivered correctly.
Retrieving intervention data Only use standard SD cards (do not use mini or micro SD cards). • To read the intervention data, use the appropriate SCHILLER software. Contact your SCHILLER representative. To retrieve the intervention data, an SD card is required. The SD card must be con- figured according to the following instructions.
6.1.5 Functional check Note: Every 6 years The replacement of the internal backup battery is advised. Should this internal backup battery not be replaced every 6 years, SCHILLER can- not ensure the proper time stamping of the intervention. Page 39...
– not placed on walls with risk of vibrations (e.g. near doors, windows, ...) Yes No Exemption from the technical safety inspection of the DefiSign Life AED if all factors are fulfilled Location: Date: Carried out by: 6.1.2...
Maintenance DefiSign Life AED User Guide Maintenance Intervals 6.1.3 Visual inspection of the device and accessories Regularly and after each use, inspect visually the device and the cables in order to detect possible mechanical damages. If you observe damages or dysfunctions which can endanger the safety of the patient or user, only use the device once it has been serviced.
LED is not blinking, the device is defective and must be repaired. In case of intensive use of the device, SCHILLER recommends that these in- spections be performed at shorter interval. The regulations in force in each country regarding inspection frequency must be observed (if shorter intervals than those recommended by SCHILLER are im- posed).
Maintenance DefiSign Life AED User Guide Cleaning and disinfection Cleaning and disinfection Shock hazard — Remove the battery before cleaning the device. This ensures that the device will not be turned on inadvertently while you are cleaning it. Risk of death! Disconnect the defibrillation pads before cleaning the device.
. A full list of all SCHILLER representatives can be found on the SCHIL-LER website (www.schiller.ch). In case of difficulty, contact SCHILLER. Our staff will be pleased to help process your order or to provide details for all SCHILLER products.
In accordance with national law, the battery must be dis- posed of at an appropriate waste disposal station or returned to SCHILLER. According to European legislation, this device is considered as electronic waste equipment.
Maintenance DefiSign Life AED Troubleshooting Troubleshooting • If it is not possible to get the device back into operating condition within a reason- able period of time, continue cardiopulmonary resuscitation until the rescue service arrives. Forced shutdown procedure • If the device cannot be switched off via normal OFF procedure (closing the cover) remove the battery and insert it again.
Page 47
Maintenance DefiSign Life AED User Guide Troubleshooting a. If 2/3 of the single measurements over a duration of 48 hours are out of the limits, the device will give a warning (Electrode and Service LEDs will light + sound alarm if configured); the status LED continues to blink.
Maintenance DefiSign Life Troubleshooting 6.5.2 Troubleshooting Forced shutdown procedure If the device cannot be switched off via normal OFF procedure remove the battery and insert it again. Problem Possible causes Remedy The Status indicator is not Replace the battery.
3 m from the device and its cables. However, there is no guarantee that no interference can occur in certain instal- ® lations. If the DefiSign Life AED causes interferences, these can be prevented by switching off the device.
Technical Data DefiSign Life AED System Specifications 7 Technical Data Unless otherwise stated, all specifications are valid at a temperature of 25 °C. System Specifications Manufactured by SCHILLER MEDICAL Device name ® DefiSign Life AED Dimensions 310 x 255 x 100 mm (h x l x w) Weight Approx.
DefiSign Life AED is not designed to be used in the presence of flammable mixtures of anaesthetic agents with air or oxygen. The SCHILLER quality management system complies in full with the international standards ISO 9001 and ISO 13485. Page 51...
Default energy settings Accuracy at 50Ω: ± 3 J or ± 15 % (the higher value is assumed) SCHILLER's customer service department can change the default energy levels to the following values: 70 – 90 –120 – 150 – 200 J (adults)...
Page 53
Technical Data DefiSign Life AED User Guide Defibrillation pulse Shock delivery • With the orange button (in semi-automatic) • Via disposable pads applied to the patient in an anterior-lateral or anterior-pos- terior position Safety discharge when: • A non shockable rhythm has been detected •...
The bandwidth of the devices that recorded the signals is larger than that of the DefiSign Life AED®. However, when the analogue signals of the database are run on the DefiSign Life AED vi a electrode connector, the DefiSign Life AED rhythm detector signal-processing characteristics are applied. Moreover these signals are of appropriate length to allow decisions to b e made by the detector system.
® The DefiSign Life AED is intended for use in the electromagnetic environment specified be- ® low. The customer or the user of the DefiSign Life AED should assure that it is used in such an environment. 7.4.1 Electromagnetic emissions...
Technical Data DefiSign Life AED Electromagnetic interferences Conformity Interference testing IEC 60601 test level Electromagnetic environment - explanations level Recommended minimum distances Portable and mobile HF telecommunication devices must keep the recommended minimum distance from the ® DefiSign Life AED and all its components, incl.
Page 57
® The DefiSign Life AED is intended to be used in electromagnetic environments in which it is possible to control radiated HF interferences. The user of the DefiSign Life ® can pre-vent electromagnetic interferences by always keeping a minimum distance between portable/mobile HF communication devices (transmitters) and the ®...
Technical Data DefiSign Life AED Literature Literature European Resuscitation Council Guidelines 2015 for Cardiopulmonary Resuscitation and Emergency Cardiovascular (2015) Care American Heart Association (2015) Guidelines 2015 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Glossary ABCD The primary ABCD A = Airways (check if airways are free)...
Date: Performed by: In case of problems, please notify your Biomedical Department , your local SCHILLER distributor , or the authorized Custom- er Service for your area Name: ..............Tel.: ..............Page 59...
Need help?
Do you have a question about the DefiSign Life and is the answer not in the manual?
Questions and answers
With the handle facing away from me the bottom red is LED flashing. The pads are due to expire in July but i have put new pads now the LED red light is still flashing. The machine is 3 years old the battery expiry date is 2031.