Battery Depleted During Use, Limited Mode (Cpr) - Schiller FRED PA-1 Series User Manual

Automated external defibrillator (aed)
Hide thumbs Also See for FRED PA-1 Series:
Table of Contents

Advertisement

3
Initial operation
3.2
Battery monitoring
1
Battery LED
Page 24
3.2.3

Battery depleted during use, limited mode (CPR)

Patient hazard — Defibrillation is no longer possible if a depleted battery is detected.
The battery needs to be replaced immediately.
If a depleted battery is detected while the device is in use, the device will prompt the
user to replace the battery and perform CPR. An audible signal is emitted. The main
status LED is off and the orange battery LED is blinking until the battery is replaced.
Depleted battery during self-test
• An audible signal is emitted, the main status LED (1) is off and the battery LED (2)
is blinking until the battery is replaced.