Defibrillation Pulse - Schiller FRED easyport plus User Manual

Automated external defibrillator (aed)
Table of Contents

Advertisement

8
Technical Data
8.3

Defibrillation pulse

Defibrillation Waveform
Accuracy for AED and manual
shock
Page 68
8.3
Defibrillation pulse
• Pulsed BTE (Biphasic Truncated Exponsential) Multipulse Biowave® with fixed
physiological optimum phase durations
• Maintains the energy delivered to the patient at an approximately constant level
with regard to patient resistance (up to 175 Ohms)
Printout: Current – left y-axis,
Deviation from the selected energy (1 till 200 J) at 25 till 175
 3 J or 15 % (the higher value is assumed)
(---
mean current calculated for each cycle)
200 Joule delivered into 25 -175 Ohm
time [ms]
50 Joule delivered into 25 -175 Ohm
time [ms]
®
®
FRED
easyport
plus
:
Rpat []