FRED easyport ® ® Automatic external Defibrillator ® ® FRED easyport (1) On-/off-button (2) Start analysis (3) Defibrillation button (4) Electrodes connector (5) Lamp ”electrodes o.k.” Short Operating Instruction Art.-No.: 2.510553 Rev.: b *2.510553*...
Page 2
Step 1 Switch on the device, apply electrodes Switch on the device by pressing the green button Apply defibrillation electrodes. Connect electrode cable to device. The yellow electrode indicator LED will go out as soon as the device is able to identify an acceptable electrode resistance.
Page 3
Step 3 Delivering the Shock Electric shock hazard! • Do not, under any circumstances, touch the patient while the shock is delivered. • Make sure that the patient is not touching any conductive objects Release the shock with the button If the shock was not successful, steps 2 to 3 are repeated once or twice.
Checking the electrodes Firmly press the defibrillation electrodes in alternation onto the patient's skin and note which electrode causes the message "Check electrodes" to disappear. Now press this electrode once again firmly against the pa- tient's skin. If the "Check electrodes" message does not disappear: –...