Technical Notification - Schiller FRED easyport plus User Manual

Automated external defibrillator (aed)
Table of Contents

Advertisement

7
Maintenance
7.6
Troubleshooting
Problem
The notification tone does
not stop.
Pads expired
Maintenance required
notification
Notification/error
Reset patient data storage
Not recording patient data
Test Not Executed, Pads
Connected
Battery Low
Charging Not Possible
Selftest failed:
– Relay test failed
– IGBT test failed
– Battery test failed
– Capacitor test failed
Notification/error
Battery empty. Shutdown In • Battery empty
Battery is very cold/hot
Shock disabled, low battery • Battery capacity to low to charge
Key stuck detected
Defibrillation failure
ECG VF/VT Detected
Page 62
Possible causes
• Selftest failed
• Battery defect.
• Device defective.
• Entered electrode expiry date ex-
pired
• Entered maintenance interval re-
minder date expired
(8.12.2 Base
Settings)
7.6.3

Technical notification

Low priority notification displayed on
Possible causes
• Failure to write intervention data
to the internal memory
• Failure to write intervention data
to the internal memory
• Connected pads during relay
selftest detected
• Battery capacity falls below 10%  If during intervention is issued make sure having a spare
• Battery capacity to low to charge
to capacitor
• one of the test below failed
• Device defective.
• Insulated Gate Bipolar Transis-
tors defective
• Battery not enough capacity or
defective
• Charging the capacitor takes to
long or is not possible
High priority technical notification displayed on
Possible causes
• Battery performance limited be-
cause temperature below/above
the allowed limits (-20/60°C)
the capacitor
• Key stuck detected
• Device failure
• You are in monitoring mode and
device detects VF/VT
Remedy
 Switch the device on, read error message, remedy the
cause and run a new selftest.
 Replace the battery.
 Have the device repaired.
 Check expiry date on the electrode packaging. Proceed
as in chapter
3.2.4
described.
 Technical safety inspection and software updates re-
quired (see
7.1 Maintenance Intervals, 3 years)
orange
background
Remedy
 After unsuccessfully retrying writing to memory the fail-
ure "Not recording patient data" appears see below.
 Get the device repaired after intervention
 Ignored notification "Disconnect Defib Pads For Test"
during the selftest. Start again selftest.
battery and replace it when possible
 Replace battery immediately
 see single tests below
 Switch the device on and check error message.
 If other error message is displayed, have the device re-
paired.
 Have the device repaired.
 Replace the battery and run again a manual self-test.
 Have the device repaired.
Remedy
 Replace battery immediately
 Replace battery with a stored battery within the temper-
ature limits. Note this message is only disaplyed when
using the rechargeable battery
 Replace battery immediately.
 Shock key was pressed during start-up of the device or
the key stucks. Cycle power off-on again and make sure
that shock key is not pressed
Have the device repaired.
 Device switches automatically to defibrillation. AED pro-
tocol starts (see
4.6 Monitoring mode)
®
®
FRED
easyport
red
background
plus