If the periodical inspection reveals problems relating to patient or caregiver security, the
technician in charge of the inspection must immediately inform the user, and add a
comment in the inspection report. The equipment should not be used until the problem is
corrected.
SCALEO Medical should be informed of the result of inspections performed on its devices.
®
Poweo
Service tool
SCALEO Medical has equipped the Poweo
boxes) with an embedded software to manage the maintenance and inspections of the Poweo
lifters. The electronic board has the following main functions:
Control of the lifter
3 jacks driving capability (one at a time).
Electrical emergency lowering.
Main actuator software safety :
o The lifter stops if the load is over loaded (max value + 10kg),
o Legs opening safety: the legs won't open or close if an obstacle is encountered.
Battery management.
Alarms. The user will be warned by a visual alarm on the remote control if the periodical
inspection date is within two months or has expired.
Defect detection:
o Main fuse out of order,
o Real time clock out of order or incorrect,
o Main actuator 97% worn.
Record capabilities
Date of first use: first power ON of the lifter. This date is used to calculate the date of first
periodical inspection.
History of :
o Transfers per weight range: [10-30kg][30kg-60kg] [60kg-110kg] [110kg 170kg-] [170kg-
SWL],
o Actuators (main and legs opening),
o Alarms,
DOMU102_EN_8_20221201_ Poweo® 215-200 user manual
®
range (except the Poweo
®
using Linak
®
or Timotion
®
control
®
patient
37