    Content WARNING AND SAFETY INSTRUCTIONS ......................3 SYMBOLS AND PICTOGRAMS ......................... 5 GENERAL INFORMATION ..........................6 Intended use ............................ 6 Contraindications ..........................7 Instructions for assembly ......................... 7 Installation of the suspension ......................9 Commissioning the Poweo ® 140 -180 Quick Up ................12 INSTRUCTIONS FOR USE ..........................
  • Page 3: Warning And Safety Instructions

    140 kg and up to 180 kg with the Poweo ® 180 Quick Up. Beyond this weight, do not use this lifter. Contact SCALEO Medical, for advice on the best lifter that satisfy your transfer needs. It is NOT recommended to use others type of suspensions from others manufacturers than the ones recommended for SCALEO Medical patient lifters.
  • Page 4 Quick Up was commercialized on Juin 4th 2020. This user manual is protected by copyright. It should not be handed over, or made available to, any unauthorized third party, without a prior written consent from SCALEO Medical DOMU125_EN_v3_20221205_Poweo® Quick Up user manual...
  • Page 5: Symbols And Pictograms

    Symbols and pictograms Symbols used in this manual: This symbol is used to point out instructions and information related to safety in the working environment, where injury may occur if the information is disregarded or ignored. Follow these instructions and be careful and attentive at all times. This symbol indicates important information regarding the use of the equipment.
  • Page 6: General Information

    General information You have recently purchased a Poweo ® 140 or a Poweo ® 180 Quick Up patient lifter. We would like to thank you for the trust you have placed in our company. The Poweo® lifter is delivered in a carton box, including the following: 1.
  • Page 7: Contraindications

    Before unpacking the patient lifter, please make sure the package is not damaged. If that is the case, immediately notify in written to the freight forwarder and contact SCALEO Medical. These instructions for assembly concern the standard version of the Poweo ®...
  • Page 8 position before operating or charging the unit. To start: Unpack the lifter by removing all protection devices and taking the lifter out of its carton box. Step 1: Install the actuator Install the actuator placing one end on the arm and one end on the mast of the lifter, thanks to a hinge.
  • Page 9: Installation Of The Suspension

    Step 3: Secure the installation with the provided screw and turn it inside the column, without tightening it. It is VERY important to properly secure the assemblage: the stability of the hoist highly depends on a strong link between the base and the mast. The screw is equipped with a thread lock to prevent the screw from loosening.
  • Page 10 Attach the suspension to the arm of the lifter by inserting the security pin as shown below: a) Insert the suspension onto the arm of the lifter b) Pass through the security pin c) Secured the assembly by introducing the washer as shown on the pictures here below: Poweo®...
  • Page 11 Do some rotational tests before using with a patient Step 8: Slide the battery into its compartment. Before using the Poweo ® make sure that the emergency button is on “OFF” position. Turn the button clockwise to release it. Position the battery on the support, and guide it for the best electrical contact between the battery and battery support contacts.
  • Page 12: Commissioning The Poweo ® 140 -180 Quick Up

    Before charging the battery, you will need to check that the plug provided fits to your local AC outlet. If the place of use was informed before ordering, SCALEO Medical will provide the right type of AC cord. Before each use check the correct installation of the lifter and of its accessories, verify proper operation before use with a patient.
  • Page 13 ® The Poweo patient lifters are designed to be used in room temperature, +10 to +40°C. This should be taken into consideration when transporting the lifter; and therefore, keeping it in a temperate part of the vehicle. Alternatively, the lifter must rest in a temperate room until the equipment has achieved proper working temperature.
  • Page 14: Instructions For Use

    Instructions for use Security devices Emergency stop: The Poweo ® patient lifters are equipped with an emergency stop button that will turn off the lifter when pressed. The emergency stop button is located on the top of the battery on the Poweo ®...
  • Page 15: Battery Management

    The opening of the legs is also secured: in the case of a contact between a leg and an obstacle, the legs will automatically stop moving. The leg opening movement stop is therefore a warning for the user: it means that the security device is operating, so the transfer of the patient is not secured.
  • Page 16 To charge the SCALEO® battery : 1. Connect the charger to the AC outlet, Emergency button 2. Place the battery in the battery charger, 3. Once the battery is detected Battery label in the charger, the LED indicator lights up. Charge status Battery status...
  • Page 17: Maintenance Of The Battery

    The external charger The external charger allows you to charge one battery while a second battery in being used. The LED indicator of the SCALEO charger displays the status of the recharge. Depending on the color of the LED the user can monitor the different status. CHARGER LED COLOR MODE Yellow -...
  • Page 18: The Controls

    warranty of batteries that would come to fail due to lack of maintenance. In case of an intensive use failing to meet the service factor of the lifter (2/10), a "battery empty" false alarm may appear on the remote control. In this case do the following: •...
  • Page 19: Transferring With The Poweo ® 140 -180 Quick Up

    Understanding the information of the remote control The maintenance LED indicator is used to indicate a possible malfunction AND the periodic inspection status. Therefore, it can display multiple consecutive status. Example: in case default AND periodical inspection date exceeded, the LED displays alternately flashing RED and solid ORANGE.
  • Page 20 Set the lifter so that the back of the patient is in contact with the back of the chair. Place the suspension above the user until the lower edge of the foam is approximately 2 cm above the thighs of the patient. Position the foams around the chest of the patient and ensure that they remain far enough below the armpits.
  • Page 21 Start the transfer. Like all patient lifters, the Poweo® is not a transport device. It must be used only for short distance transfers and is not intended to replace wheelchairs or other types of transport. Position the patient as low as possible during the transfer, in order to lower the centre of gravity and ensure maximum stability.
  • Page 22 1. Undress the patient. 2. Position the suspension around the patient as previously described. 3. Lift the patient above the bathtub using the patient's legs to deposit him into the bathtub. 4. Maintain the patient slightly lifted so as the foams can hold him.
  • Page 23: Cleaning And Disinfecting

    Cleaning and disinfecting The following program for cleaning and disinfection has been established in compliance with the “Guide to good practices for disinfection of medical devices” edited by the French council of public Health. Device classification : Not critical Infectious hazard: low risk Level of treatment required: Low-level disinfection This procedure is relevant to any medical device of the described category : apply the items when relevant...
  • Page 24 3. Disinfect: destroy or inactivate micro-organisms. 4. Dry: protect equipment from contamination once disinfected. 5. Store: keep the disinfected devices clean. Products that should be avoided for cleaning and disinfecting: • Peracetic acid at concentrations> 10%, • Organic solvents like halogenated hydrocarbons / aromatic or acetone, •...
  • Page 25: Periodic Inspection And Preventive Maintenance

    Periodic inspection and preventive maintenance The aim of a periodic control visit is to control the safety and the maintenance issues on the equipment. This control, as well as the preventive maintenance must be performed regularly every year for the patient lifter and twice a year for the slings – according to annex B of the EN ISO 10535 standard and the recommendations from health authorities.
  • Page 26 Fixation of the base with the column Cleaning and lubrication of wheels (front & back) Electric emergency lowering Electric opening of the base Foot parallelism (closed position) Column: Check the wear level of the joints Directional handles Check the actuator hooks Actuator (wear level) Battery holder Power contacts...
  • Page 27: Poweo ® Service Tool

    The equipment should not be used until the problem is corrected. SCALEO Medical should be informed of the result of inspections performed on his devices. Poweo ®...
  • Page 28 • Defect detection: o Main fuse out of order, o Real time clock out of order or incorrect, o Main actuator 97% worn. Record capabilities • Date of first use: first power ON of the lifter. This date is used to calculate the date of first periodical inspection.
  • Page 29 Version : Distributors Features PC software Card interface Main fuse test Real time clock repair Main actuator control Periodical inspection period setting Setting the date of periodical inspection at current date Disable the periodical inspection alarm Real time setting of the lifter Control of the lifter without the remote control ®...
  • Page 30: Technical Specifications

    The lifetime of the Poweo® patient lifters is 10 years or 30,000 cycles, provided that the recommended periodic inspections and preventive maintenance are properly made. SCALEO Medical lifters are compliant with the 2011/65/EU European directive relating to reduction of hazardous substances. An EU declaration of conformity can be provided on request.
  • Page 31 Poweo® 180 Quick Up suspension DOMU125_EN_v3_20221205_Poweo ® Quick Up user manual...
  • Page 32: Technical Specifications Of The Poweo 140 - 180 Quick Up

    Technical specifications of the Poweo ® 140 – 180 Quick Up Poweo® 180 Quick Up Poweo® 140 Quick Up Specifications in mm in mm Highest position 1387 1380 Lowest position Hoisting range 1196 1106 External width (legs opened) 1391 1199 External width (legs closed) Maximum length (legs closed) 1555...
  • Page 33: Technical Specifications Of The Accessories

    Other specifications: Poweo® 140 Quick Up Poweo® 180 Quick Up Maximum load 140 kg 180 kg Lifts with a 100% charged battery ≥ 80 ≥ 50 Classification Class I medical devices (93/42/EEC Directive) Materials Steel and aluminum Lifting actuator 24V – 10A Front wheels 2 x Ø...
  • Page 34 intensity, short-circuit ... - Built-in electronic protection against overheating during charging Technical specifications of the S1750001 battery charger: The provided charger is able to work with any AC power value: voltage from 100V to 240V, frequency 50Hz or 60Hz. The only adaptation needed is the AC cord. Input AC voltage 100-240V AC Input AC frequency...
  • Page 35: Electromagnetic Compatibility (E.m.c)

    Electromagnetic compatibility (E.M.C) • ® The Poweo patient lifters require special care regarding the E.M.C. They must be installed and commissioned by the EMC information provided in this user manual. • The RF communication devices and portable phones may affect the Poweo ®...
  • Page 36 The Poweo® is designed to be used in the electromagnetic environment specified below. The user should ensure it is used in such an environment. INMMUNITY test Test level - IEC 60601 Level of Electromagnetic environment guidelines compliance Electrostatic ± 8 kV contact ±...
  • Page 37: Complementary Safety Measures

    2. If the lifter breaks down or has a malfunction, do not try to repair it yourself. You should seek the help of a qualified technician, an authorized distributor, or the customer service department of SCALEO Medical. 3. Always demand original authorized spare parts.
  • Page 38 ® Injury or damage to the equipment may occur if the Poweo is not properly used, more precisely if: • The protective covers of the chassis are taken off by non-qualified personnel, • A regular maintenance is not set up, •...
  • Page 39: Warranty

    Warranty The Poweo ® 140 and 180 Quick Up patient lifters have a four (4) years warranty, except for the battery that is guaranteed one (1) year. • Any intervention on the machine contrary to this manual will entirely void the warranty. •...
  • Page 40: Liability

    Liability SCALEO Medical assumes no liability for any injury or harm and consequences thereof directly or indirectly caused to operators, patients, or any third party in the following cases: • Non-compliance of the instructions and recommendations supplied within the present user manual, •...
  • Page 41: End Of Life And Recycle

    End of life and recycle SCALEO Medical lifters and accessories are at least 90 % recyclable. The end of life disposal of the device must be done respecting the recycling circuit and local relevant regulation. This particularly concerns the electrical and electronic equipment (WEEE) and the lead battery (B/A).
  • Page 42: Ec Declaration Of Conformity

    EC Declaration of conformity DOMU125_EN_v3_20221205_Poweo® Quick Up user manual...
  • Page 43 DOMU125_EN_v3_20221205_Poweo ® Quick Up user manual...
  • Page 44: Contact Information

    Contact information The information and advice provided in this user manual are also intended to inform you of the importance of paying attention to safety measures and the necessity of regular maintenance. This is a matter of responsibility for your establishment, and a question of safety for your patients and staff.

This manual is also suitable for:

Poweo 180 quick up

Table of Contents