bolts, one at each side. Follow the same procedure for the other end.
The bolts are coated by the "LOCTITE 243" to avoid the lost of screws during the lifting operating
mode.
The "LOCTITE 243" is a medium thread lock coated on the screws for the future installation at
the customers facilities.
Tighten with a hexagonal wrench or the provided special wrench, without using too much force: the
screw head would be deteriorated and it would be impossible to dismount it later.
These operations are very important for the security of the patient and caregivers.
Without the security pins, the column or the arm may fall and provoke serious injuries.
You must ensure that these pins are properly secured and perform regular
inspections: screws must ALWAYS be present and in good condition. A label near the
pins allows a quick identification for inspection.
Step 2: Install the mast on the base.
Step 3: Secure the installation with the provided screw and turn it inside
the column, without tightening it.
It is VERY important to properly secure the assemblage: the
stability of the hoist highly depends on a strong link between
the base and the mast.
The screw is equipped with a thread lock to prevent the screw
from loosening. Install the screw on two or three tread
without completely tightening.