Commissioning The Poweo 200 And Poweo 215 - Scaleo medical Poweo 200 User Manual

Patient lifters
Table of Contents

Advertisement

Position the battery into its compartment, and guide it for the best electrical contact
between the battery and battery support contacts.
Step 7: Connect the remote control.
Connect the remote control with the plug and make sure that the wire
connection is aligned with the mast remote control port. Once aligned, press
on the wire to insert the plug.
Step 8: Test the lifter.
Try lifting and lowering, opening and closing the legs to make sure that everything is in order.
Step 9: Remove the battery and recharge it.
Before first use, the battery must imperatively be 100% charged. Please refer to the "Battery
management" section for this operation. Your lifter will be ready for use as soon as the battery is
charged.
Whenever possible, keep the carton box and packing accessories for later transportations.
Commissioning the Poweo
SCALEO Medical provides an universal power supply adapted to any voltage/frequency
value. Before charging the battery, please check that the plug provided fits to your local AC
outlet.
If the place of use was informed before ordering, SCALEO Medical will provide the right type of
AC cord.
Before each use check the correct installation of the lifter and its accessories, verify proper
operation before using with a patient.
The control points presented below only take a few minutes and allow you to ensure a safe use of
®
your Poweo
patient lifter: