and add a comment in the inspection report. The equipment should not be used until the problem
is corrected.
SCALEO Medical should be informed of the result of inspections performed on his devices.
Poweo
®
Service tool
SCALEO Medical has equipped the Poweo
maintenance and inspections of the Poweo
main functions:
Control of the lifter
3 jacks driving capability (one at a time).
Electrical emergency lowering.
Main actuator software safety :
o The lifter stops if the load is over loaded (max value + 10kg),
o Legs opening safety: the legs won't open or close if an obstacle is met.
Battery management.
Alarms. The user will be warned by a visual alarm on the remote control if the periodical
inspection date is within two months or has expired.
Defect detection:
o Main fuse out of order,
o Real time clock out of order or incorrect,
o Main actuator 97% worn.
Record capabilities
Date of first use: first power ON of the lifter. This date is used to calculate the date of first
periodical inspection.
History of :
o Transfers per weight range: [10-30kg][30kg-60kg] [60kg-110kg] [110kg-170kg]
[170kg- SWL],
o Actuators (main and legs opening),
o Alarms,
33
®
300 with an embedded software to manage the
®
patient lifters. The electronic board has the following
DOMU124_EN_v2_20221206_Poweo® 300 User manual