Never recharge batteries in extreme temperatures. If the temperature is too high, the
battery charger may stop functioning for security reasons. If this occurs, place the battery
and the charger in a cool place, wait a few minutes so the security device let the charge
begin, and try again to recharge the battery.
If you plan to not use your lifter for a long time (≥ 1 month), proceed as follows to keep
your batteries in good conditions:
Fully recharge your battery,
Push the emergency button to OFF position,
Remove the battery from its compartment,
Store your battery charged in a cool, dry, well ventilated area,
Check periodically (every month) the status of charge of the batteries: even if they
are disconnected, batteries tend to discharge.
Compliance with these precautions will help you keep your batteries longer and use your
equipment in the best conditions. Failure to follow these instructions will void the warranty of
batteries that would come to fail due to lack of maintenance.
In case of an intensive use failing to meet the service factor of the lifter (2/10), a "battery
empty" false alarm may appear on the remote control. In this case do the following:
Remove the battery from its compartment,
Wait 10 seconds,
Put the battery back on its compartment.
This operation allows the electronics to reset and cancels the "false" alarm.

The controls

Remote controls for Poweo
The Poweo® patient lifter equipped with the SCALEO
remote control and one fixed control panel installed on the column of the patient lifter. The user can
DOMU102_EN_8_20221201_ Poweo® 215-200 user manual
®
®
215 and Poweo
200 equipped with SCALEO electronics
®
power pack battery has two controls: one
19