Scaleo medical Poweo 200 User Manual page 47

Patient lifters
Table of Contents

Advertisement

INMMUNITY
Test level - IEC 60601
test
Electrostatic
± 8 kV contact
discharges
± 15 kV air
(E.S.D.)
IEC 61000-4-2
Transitional
± 2 kV for power supply
burst
lines
IEC 61000-4-4
± 1 kV for input/output
lines
Transitional
± 1 kV between phases
surge
± 2 kV line to ground
IEC 61000-4-5
Voltage dips,
<5% UT (> 95% dip in
short
UT)
interruptions
for 0.5 cycle 40% UT
and voltage
(60% dip in UT) for 5
variations on the
cycles
input lines
70% UT (30% dip in UT)
Power
for 25 cycles <5% UT (
IEC 61000-4-11
>95% dip in UT) 5 s
Magnetic field
30 A/m
@ 50/60 Hz
IEC 61000-4-8
NOTE U
is the AC mains voltage prior to application of the test level.
T
The Poweo® is designed to be used in the electromagnetic environment specified below. The user of the Poweo® must ensure it is used in such an
environment.
Conducted
3 Vrmsf
RF
150 kHz to 80
disturbances
MHz
IEC 61000-4-
6
3V/m 80MHz
Radiated RF
to 2.7GHz with
disturbances
discreet
IEC 61000-4-
frequencies
3
385, 450, 710,
745, 780, 810,
870, 930, 1720,
5240, 5500,
5785, 1845,
1970, 2450
with level
between 9V/m
and 28V/m
DOMU102_EN_8_20221201_ Poweo® 215-200 user manual
Level of compliance
± 8 kV contact
± 15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
Directives and manufacturer's declaration - electromagnetic immunity
Portable and mobile RF communications should not be used closer to any part of the Poweo®, including
cables, than the recommended separation distance calculated from the equation applicable to the
frequency of transmitter.
Recommended separation distance
Non
Non applicable
applicable
3 V/m
d= 0,35√P 80MHz to 800MHz
d= 0,7√P de 800MHz to 2,5GHz
Where P is the maximum power output of the transmitter in watts (W), according to the transmitter
manufacturer and d is the separation distance in meters (m). The field strengths from fixed RF
transmitters, as determined by an electromagnetic investigation (a) should be below the level of
compliance in each frequency range. (b) Interference may occur in the vicinity of the equipment marked
with the following symbol:
Electromagnetic environment guidelines
Floor materials should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, should the relative humidity be at least 30%.
Not applicable
Not applicable
Not applicable
Quality of power supply network should be of a typical commercial or hospital
environment. If the user of Poweo® requires continued operation during network
outages of power, it is recommended to feed the Poweo® from an uninterruptible
power supply or battery.
47