Scaleo medical Poweo 200 User Manual page 21

Patient lifters
Table of Contents

Advertisement

Understanding the information of the remote control
The maintenance LED indicator is used to warn a possible malfunction AND the periodic
inspection status. Therefore, it can display multiple consecutive status. Example: in case of
default AND periodical inspection date exceeded, the LED displays alternately flashing RED
and solid ORANGE.
Managing the maintenance LED status is easy:
Continuous solid light = default.
The light is continuous and the lifter is not operating = main fuse replacement is necessary.
Flashing light = periodical inspection management.
According to the EN 10535 standards, the period between periodical inspections is by established by
default to 12 months. The periodical inspection time setting can be changed using the Poweo® service
tool software.
BATTERY LED
– Solid green = more than 10 cycles are still
possible.
– Solid red: = less than 10 cycles left before
battery empty status.
– Flashing red: battery alarm (only one cycle
left before the battery is completely empty).
Remote controls for Poweo
Two remote controls are available: the standard remote control for models with a control box without
display, and the remote control for models equipped with a control box with display.
DOMU102_EN_8_20221201_ Poweo® 215-200 user manual
– Solid orange: default. Use the Poweo
tool software or call us for maintenance.
– Flashing orange: periodic inspection must be
performed within 2 months.
– Flashing red: periodic inspection has not been
performed on time.
®
215 equipped with Linak
MAINTENANCE LED
®
electronics
®
service
21