Emergency stop
Labelling:
Control label
Manufacturer or distributor logo
Number of completed actions
Name of technician:
If the periodical inspection reveals problems relating to the security of the patient or
the caregiver, the technician in charge of the inspection must immediately inform the
user, and add a comment in the inspection report. The equipment should not be used until the
problem is corrected.
SCALEO Medical should be informed of the result of inspections performed on his devices.
Poweo
®
Service tool
SCALEO Medical has equipped the Poweo
control boxes) with an embedded software to manage the maintenance and inspections of the
Poweo
®
patient lifters.
The electronic board has the following main functions:
Control of the lifter
3 jacks driving capability (one at a time).
Electrical emergency lowering.
Main actuator software safety :
o The lifter stops if the load is over loaded (max value + 10kg),
o Legs opening safety : the legs won't open or close if an obstacle is met.
Battery management.
Alarms. The user will be warned by a visual alarm on the remote control if the periodical
inspection date is within two months or has expired.
DOMU125_EN_v3_20221205_Poweo ® Quick Up user manual
Date :
Signature :
®
range (except the Poweo
®
using Linak
®
or Timotion
®
27