Cleaning and disinfecting
The following program for cleaning and disinfection has been established in compliance with the
"Guide to good practices for disinfection of medical devices" edited by the French council of public
Health.
Device classification: Not critical
Infectious hazard: low risk
Level of treatment required: Low-level disinfection
This procedure is relevant to any medical device of the described category: apply the items when
relevant
General guidelines
•
Always respect the general hygiene precautions (gloves, goggles, protective clothes and mask).
•
Disinfect your hands after removing your protection gloves.
•
Respect the expiration dates of products used.
•
Respect the dilutions.
•
Never mix different cleaning or disinfection products.
•
In case of contact with eyes, rinse thoroughly with water; call a doctor in case of persistent
pain.
•
Never perform a cleaning or disinfection operation when the device is powered: disconnect
the device from the AC outlet.
Recommended procedures for cleaning and disinfecting
•
The device should be thoroughly cleaned and disinfected at least once a month.
•
The device should be systematically and thoroughly cleaned and disinfected between 2
patients.
1. Prepare the equipment for disinfection: Purpose: Makes it easier to clean, lower the level of
contamination, protects the staff and the environment.
2. Clean: eliminate dirt.
3. Disinfect: destroy or inactivate micro-organisms.
4. Dry: protect equipment from contamination once disinfected.
5. Store: keep the disinfected devices clean.