Make sure that no part is damaged or show signs of wear,
Check the tightness of all the safety screws and pins,
Inspect accessible screws and make sure none are loose,
Check the lights on the remote control: battery indicator should be solid green;
maintenance indicator should be off,
Check the features: raise and lower the arm, leg opening and emergency stop.
In case of malfunction or doubt about the proper operation, do not use the Poweo®
patient lifter. Inform your technical service, your distributor, or contact us at the
address on the back of this manual.
The Poweo
®
patient lifters are designed to be used in room temperature, +10 to +40°C. This
should be taken into consideration when transporting the lifter; and therefore, keeping it in a
temperate part of the vehicle. Alternatively, the lifter must rest in a temperate room until the
equipment has achieved proper working temperature.