Electromagnetic Compatibility (E.m.c.) - Scaleo medical Poweo 200 User Manual

Patient lifters
Table of Contents

Advertisement

Electromagnetic compatibility (E.M.C.)

The Poweo
®
patient lifters require special care regarding the E.M.C. They must be installed and
commissioned considering the EMC information provided in this user manual.
The RF communication devices and portable phones may affect the Poweo
The use of accessories, transducers and cables other than those specified, except transducers
and cables sold by the manufacturer as replacement parts for internal components, can result
in increased emissions or decreased immunity of the device.
It is appropriate that the Poweo
them, if it is not possible to do otherwise, the Poweo
operation in the configuration in which it is used.
In terms of the standard NF EN 60601-1-2: 2007, essential performance of the Poweo
never perform unwanted movements.
The Poweo® is designed to be used in the electromagnetic environment specified below. The user of the Poweo® must make sure to use it in such
Emission test
Conformity
RF Emissions
Group 1
CISPR 11
RF Emissions
Class B
CISPR 11
Harmonic Emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations /
Not applicable
Flicker IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment and the Poweo®
The Poweo® is designed to be used in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the Poweo® can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the Poweo®, as recommended below, according the maximum transmission power of the communication device.
Maximum rated output
power of the transmitter
W
0,01
0,1
1
10
100
The Poweo® is designed to be used in the electromagnetic environment specified below. The user should ensure it is used in such an environment.
46
®
patient lifter is not used next to other devices or stacked with
Directives and manufacturer's declaration - Electromagnetic emissions
environment.
The Poweo® uses RF energy only for its internal functions. Therefore, its RF emissions are very low and are
not likely to cause interference in nearby electronic devices.
The Poweo® is suitable for use in all premises, including domestic premises and those directly connected
to the public power network supplying domestic premises.
Separation distance depending on the frequency of the transmitter
from 150 kHz to 80 MHz
80MHz to 800MHz
Non applicable
Non applicable
Non applicable
Non applicable
Non applicable
Non applicable
Directives and manufacturer's declaration - electromagnetic immunity
DOMU102_EN_8_20221201_ Poweo® 215-200 user manual
®
must be monitored to verify normal
Electromagnetic environment - Guidelines
m
d= 0,35√P
0,04
0,11
0,35
1,1
3,5
®
lifters.
®
is to
de 800MHz to 2,5GHz
d= 0,7√P
0,07
0,22
0,7
2,2
7