Thermo Scientific Barnstead Pacific TII 7 Operating Instructions Manual page 91

Table of Contents

Advertisement

Step
Action
3
Remove the tubing on the left side of storage tank
from "Normal Use" and reconnect into "Emergency
supply. When you hear an audible click you can be
sure that the quick connector have been inserted
correctly.
4
Turn on system and water supply.
5
Check the water pressure on the pressure gauge on
the right side of storage tank. Adjust the pressure to 2
bar/ 29 psi if necessary, by locating the pressure
reducer, inside of back panel and pull out the
adjusting knob.
Turning the knob clockwise to increase the pressure
and counter clockwise to reduce the pressure. After
adjusting the pressure, press in the adjusting knob to
lock it.
Do not forget to re-adjust the pressure when you are
returning the Pacific AFT system to the normal use
Thermo Scientific
Figure
Dispensing
valve
ON/OFF
switch
Adjustment
knob
Increase
pressure
13
Maintenance
Emergency
supply
Normal use
Storage tank
Pressure gauge
Reduce
pressure
Pacific Water Purification Systems
81