Thermo Scientific Barnstead Pacific TII 7 Operating Instructions Manual page 87

Table of Contents

Advertisement

Step
Action
1
Switch the Pacific System off and shut off the supply
of feedwater.
2
Remove the cartridge cover and take off the DI-
cartridge.
3
Unscrew the bracket from the mounting plate and
take it up over the UV lamp cable.
4
Draw the UV lamp housing slightly to the front (see
red arrow) to the front and take the plug off of the UV
lamp.
5
Now carefully draw the UV lamp upwards while
lightly turning it clockwise. During the replacement
of a UV lamp, great care must be taken to avoid
touching the glass of the UV lamp with fingers, to
avoid dirtying of the lamp which would impair the
functioning of it.
We recommend that clean gloves to be worn.
Thermo Scientific
Always wear safety gloves when changing the UV lamp, in order to prevent that your skin
comes in contact with the UV lamp glass.
Wear directly a breathing protector when you are seeing that the glass of the UV lamp is
broken and ventilate the room well.
Contact your local Service organization to proceed as the proper disposal of the used UV
lamp. The Hg content in the UV lamp is so low so that no damage to the environment can
arise.
(applicable only for systems with UV lamp)
Figure
ON/OFF
switch
See chapter
"Replacing the DI cartridge (only Pacific TII/ TII
UV, AFT)"
on
page
67.
Bracket
Plug of the UV
lamp
UV lamp
UV lamp
housing
13
Maintenance
UV lamp cable
UV lamp
housing
UV lamp
housing
Pacific Water Purification Systems