Step
Action
7
Put the cartridge cover back on (see under section
"Replacing the DI cartridge (only Pacific TII/ TII UV,
AFT)"
on
page
67), re-open the feed water supply
and switch the system on again.
8
Push the menu button until "Enter code" is
displayed.
The Code to do this transaction please refer from the
Code table under section
page
57. You need a level 3 code.
a.) After entering the code push the Menu and UV
button simultaneously. The display shows UV
Menu.
b.) Push the Menu button repeatedly until new UV
lamp appears and press enter to confirm.
The system sets the operating hours counter of the
UV lamp back and save the new values by an
automatically calibration.
The UV lamp must be switched on (production
mode).
Thermo Scientific
"Unlocking the System"
on
Figure
a.
b.
13
Maintenance
UV Menu
UV Menu
new UV lamp
press enter
Pacific Water Purification Systems
79