Maintenance
Step
Action
6
Ensure that the position of the sealing ring (flat
o-ring at the top of the connection nut) is correct as
you put in the new UV lamp, otherwise you will have
a leak. The sealing ring must be seat in the rabbet of
the connecting nut (see picture a and b).
Carefully introduce the new UV lamp under a slight
turning motion like before but in the anti-clockwise
direction (see picture c). Attach the plug into the
lamp and push the housing back to the system.
Once it is in place, re-mount the bracket holding the
UV housing onto the system's the mounting plate.