Thermo Scientific Barnstead Pacific TII 7 Operating Instructions Manual page 83

Table of Contents

Advertisement

Step
Action
1
For all systems:
Drain tank from front spigot until system says
Production. Turn system off.
2
For Pacific TII and AFT systems using the tank with
recirculation pump only, install disinfection adapter,
50133431 in place of DI cartridge. When you hear an
audible click you can be sure that the quick
connectors have been inserted correctly. Discard used
DI cartridge.
Pacific TII and AFT systems using tank with
recirculation pump:
Do not allow system to recirculate with DI cartridge in
place. If disinfection adapter is not available, keep
system off during disinfection process.
Pacific TII systems using tank without pump, Pacific
RO, skip to step 3.
3
Turn on the system. Program the Pacific to fill the tank
completely by adjusting the float switch to turn off at
100%. See
"Adjusting the settings of the Float
Switch"
on
page
Pacific TII using the tank with recirculation pump only:
Adjust recirculation time in OEM menu to 30 minutes.
See
"Recirculation Time"
Thermo Scientific
If Pacific is feeding Type I system or other equipment, turn off and do not dispense water
from Type I system or other equipment during disinfection process to avoid bleach solution
from entering those systems.
63.
on
page
60.
Figure
front tank
spigot
ON/OFF
switch
Quick connectors DI cartridge
Production 55%
13,7 MΩ·cm
15.5 °C UV/TC
13
Maintenance
Disinfection adapter
50133431
Stand-by 100%
13,7 MΩ·cm
15.5 °C UV/TC
Pacific Water Purification Systems
73