Installation - Agilent Technologies G1992A User Manual

Nanospray ion source
Table of Contents

Advertisement

1

Installation

Installation
The source can be configured in three ways:
16
Orthogonal Orientation
above. The needle is at right angles to the MS
capillary. Adjustment is along Z and Y axes.
Camera is to the left, viewing the spray
orthogonally.
Inline Horizontal Orientation
from the front. The needle is parallel with the
capillary. Needle movement is along Z and Y.
The camera is mounted to the left, viewing
the needle movement orthogonally. In this
orientation, the camera plate is located over
the top hole of the chamber. The camera plug
covers the opening.
Inline Sideways Orientation
from the front. The needle is parallel to the
capillary. Needle movement is forward and
side-to-side (Z-X). The camera adapter plate
is located over the top hole of the chamber,
and the camera is mounted above, viewing
the needle movement orthogonally. The
camera plug is inserted into the hole vacated
by the camera.
Nanospray Ion Source User Guide
Sample enters from
Sample enters
Sample enters