3
Installing the Module

Installation of Heater Devices

Installation of Heater Devices

Availability of heater devices
Additional to the built-in heat exchangers, external heater devices with a small
internal volume for low dispersion applications are available. The delivered
low dispersion heat exchanger should always be used to minimize dispersion.
It can be mounted in different locations. Additionally available valve kits will
have as many additional low dispersion heat exchangers as columns are
supported.
Depending on the application, these heater devices can be fixed at various places.
N O T E
Information about the usage of these heater devices can be found in the Agilent 1290
Infinity LC System - System Manual and Quick Reference (G4220-90301).
Figure 15
If the additional heat exchangers are used as shown in this figure, the column identification
N O T E
system cannot be used. If the column identification system is required, fix the heater
devices in the upper or lower locations or right/left of the current location.
56
Installation points for heat exchanger