To Clean The Wick And Wick Cover - Agilent Technologies G1992A User Manual

Nanospray ion source
Table of Contents

Advertisement

2
Mass Reference Solution

To clean the wick and wick cover

1 Use a T-6 driver to remove the four screws from the wick cover.
2 Use tweezers to remove the wick.
3 Soak the wick in a suitable organic solvent such as acetonitrile or LC
4 Allow the wick to dry completely.
5 Clean the wick cavity with IPA or LC flushing solvent.
6 Use tweezers to reinsert the wick into the slot of the top plate, or press the
7 Remove any stray fibers with tweezers, or blow them off with compressed
52
flushing solvent to clean the wick.
wick firmly into place while wearing clean gloves.
air or nitrogen, and reinstall the wick cover.
A replacement wick is available if needed (part number G1988-20021).
Nanospray Ion Source User Guide