DC Gain Accuracy Test

When using the 50 Ω channel inputs, you must ensure that the input voltage to the
CAUTION
oscilloscope never exceeds ±5 V.
Let the oscilloscope warm up before testing
N OTE
The oscilloscope under test and other test equipment must be warmed up (with the
oscilloscope application running) for at least 30 minutes prior to the start of any performance
test.
Specifications
DC Gain
Accuracy
50 Ω Input Impedance: Full scale is defined as 8 vertical divisions. Magnification is used below
10 mV/div. Below 10 mV full scale is defined as 80 mV. The major scale settings are 5 mV, 10 mV,
20 mV, 50 mV, 100 mV, 200 mV, 500 mV and 1 V.
1 MΩ Input Impedance: Full scale is defined as 8 vertical divisions. The major scale settings are
5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, and 5 V.
Equipment
Required
Description
Digital Multimeter
Cable Assembly (2
required)
DC Power Supply
Adapter
2 Adapters
Infiniium 9000 Series Oscilloscopes Service Guide
±2% of full scale at full resolution channel scale (±5 °C from calibration
temperature)
Critical Specifications
DC voltage measurement accuracy better than
±0.1% of reading
50 Ω characteristic impedance, BNC (m)
connectors
40 V
BNC Tee (m)(f)(f)
2 BNC (f) to dual banana
3
Testing Performance
Recommended
Model/Part Numbers
Keysight 34401A or
Keysight 3458A
Keysight 8120-1840
Keysight 6614C
Keysight 1250-0781
Keysight 1251-2277
35