Testing Performance

Analog Bandwidth - Maximum Frequency Check

Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
CAUTION
Let the oscilloscope warm up before testing
N OTE
The oscilloscope under test and other test equipment must be warmed up (with the
oscilloscope application running) for at least 30 minutes prior to the start of any performance
test.
Specifications
Analog Bandwidth (-3 dB)
DSO/MSO9404A
DSO/MSO9254A
DSO/MSO9104A
DSO/MSO9064A
Analog Bandwidth (-4 dB)
DSO9204H
DSO9104H
DSO9054H
DSO9024H
40
All vertical scale settings used for the 1 MΩ test (including the 5 mV/div
setting) and all 50 Ω test vertical scale settings used except for the 5 mV/div
setting should then use the following equation:
4 GHz
2.5 GHz
1 GHz
600 MHz
2 GHz
1 GHz
500 MHz
250 MHz