DC Gain Accuracy Test

C A U T I O N
Ensure that the input voltage to the oscilloscope never exceeds ±5 V.
Let the oscilloscope warm up before testing.
The oscilloscope under test must be warmed up (with the oscilloscope application running) for at least
30 minutes prior to the start of any performance test.
Specifications
DC Gain Accuracy
Equipment Required
Description
Digital Multimeter
Cable Assembly
(2 required)
Adapter
Adapter
Connector Saver
Adapter
(2 required)
±2% of full scale at full resolution channel scale
Full scale is defined as 8 vertical divisions. Magnification is used below 10 mV/div. Below 10
mV/div full scale is defined as 80 mV. The major scale settings are 10 mV,
20 mV, 50 mV, 100 mV, 200 mV, 500 mV and 1 V.
Critical Specifications
DC voltage measurement accuracy better than ±0.1% of reading
50Ω characteristic impedance, BNC (m) connectors
BNC Tee (m)(f)(f)
BNC (f) to dual banana
3.5 mm (f)-(f), shipped with each MSO/DSO/DSAX 90000A and
DSO/DSA93004L Series oscilloscope
BNC (f) to SMA (m) adapter
Chapter 3: Testing Performance
DC Gain Accuracy Test
Recommended
Model/
Part Numbers
Keysight 34401A or
Keysight 3458A
Keysight 8120-1840
Keysight 1250-0781
Keysight 1251-2277
Keysight 5061-5311
(used with 20, 25, 28, 30,
and 32 GHz oscilloscope
models)
Keysight 1250-3758
(used with 13 GHz and 16
GHz oscilloscope
models)
Keysight 1250-1200