Performance Test Record - Keysight Infiniium 9000 Series Service Manual

Hide thumbs Also See for Infiniium 9000 Series:
Table of Contents

Advertisement

3
Testing Performance

Performance Test Record

Keysight Technologies
Model Number:
Serial Number:
Recommended Test Interval—1 Year/2000 hours
Recommended next test date:
Table 4 Input Impedance Test
Input
Vertical
Resistance
Scale
50 Ω
5 mV/div
50 Ω
10 mV/div
50 Ω
20 mV/div
50 Ω
50 mV/div
50 Ω
100 mV/div
50 Ω
200 mV/div
50 Ω
500 mV/div
50 Ω
1 V/div
1 MΩ
5 mV/div
1 MΩ
10 mV/div
1 MΩ
20 mV/div
1 MΩ
50 mV/div
1 MΩ
100 mV/div
1 MΩ
200 mV/div
1 MΩ
500 mV/div
1 MΩ
1 V/div
1 MΩ
2 V/div
1 MΩ
5 V/div
52
Channel 1
Channel 2
Keysight Infiniium 9000 Series Oscilloscopes
Tested by:
Work Order No.:
Date:
Ambient temperature:
Channel 3
Channel 4
Infiniium 9000 Series Oscilloscopes Service Guide
Limits
48.75 Ω to 51.25 Ω
48.75 Ω to 51.25 Ω
48.75 Ω to 51.25 Ω
48.75 Ω to 51.25 Ω
48.75 Ω to 51.25 Ω
48.75 Ω to 51.25 Ω
48.75 Ω to 51.25 Ω
48.75 Ω to 51.25 Ω
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ
990 kΩ to 1.010 MΩ