5
Replacing Assemblies
To remove and replace the front panel bezel, front panel keyboard,
touch screen, and Autoprobe board
1
2
3
100
Disconnect the power cable and remove the handle and rear cover as described
above in the section on removing the handle and rear cover.
Position the oscilloscope upright and disengage the series of tabs around the
outside of the front bezel as shown in the following figure.
With the tabs disengaged, pull the front bezel off as shown in the following
figure.
Tabs on right side of
instrument
Infiniium 9000 Series Oscilloscopes Service Guide