7.2 Technical data

Thermocouple
input
Resistance
thermometer
input
74
Fe-Con L
Fe-Con J EN 60584
Cu-Con U
Cu-Con T EN 60584
NiCr-Ni K EN 60584
NiCr-CuNi E EN 60584
NiCrSi-NiSi N E 60584
Pt10Rh-Pt S EN 60584
Pt13Rh-Pt R EN 60584
Pt30Rh-Pt6Rh B
EN 60584
Shortest measurement span
Range start/end
Cold junction
Cold junction accuracy
(internal)
Cold junction temperature
(external)
Measurement time
Input filter
Features
1. The linearisation accuracy refers to the maximum measurement span.
The linearisation accuracy is reduced for shorter spans.
Pt 100 EN 60751
Pt 100 JIS
Pt 500 DIN
Pt 1000 DIN
Ni 100
Connection type
Shortest measurement span
Probe lead resistance
Measurement range
-200 to
+900°C
-210 to +1200°C
-200 to
+600°C
-270 to
+400°C
-270 to +1372°C
-270 to +1000°C
-270 to +1300°C
-50 to +1768°C
-50 to +1768°C
0 to
1820°C
Types L, J, U, T, K, E, N:
Types S, R, B:
freely programmable within the limits, in 0.1°C steps
Pt100 internal or thermostat as external constant
-50 to +100°C, adjustable through setup software
for 3 channels < 2sec; for 6 channels < 4sec
2nd. order digital filter; filter constant adjustable from 0 to 50.0sec
also programmable in °F; customer-specific linearisation
Connection
Range
2/3-wire
-200 to +250°C
2/3-wire
-200 to +850°C
4-wire
-200 to +250°C
4-wire
-200 to +850°C
2/3-wire
-200 to +260°C
2/3-wire
-200 to +649°C
4-wire
-200 to +260°C
4-wire
-200 to +649°C
2/3-wire
-200 to +150°C
2/3-wire
-200 to +850°C
4-wire
-200 to +150°C
4-wire
-200 to +850°C
2/3-wire
-200 to +250°C
2/3-wire
-200 to +850°C
4-wire
-200 to +250°C
4-wire
-200 to +850°C
2/3-wire
-60 to +125°C
2/3-wire
-60 to +180°C
4-wire
-60 to +125°C
4-wire
-60 to +180°C
2-, 3- or 4-wire circuit
max. 30 Ω per core for 4-wire circuit
max. 20Ω per core for 2- and 3-wire circuit
for Pt100 up to 260°C max. 10 Ω per core in
2- and 3-wire circuit
Linearisation accuracy
±0.2%
±0.2% above -200°C
±0.3%
±0.5% above -200°C
±0.2% above -150°C
±0.2% above -200°C
±0.2% above -150°C
±0.5% above 0°C
±0.5% above 0°C
±0.5% above 500°C
100°C
500°C
± 1°C
Linearisation
accuracy
±0.6°C
±1.0°C
±0.5°C
±0.8°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
±0.6°C
±1.0°C
±0.5°C
±0.8°C
15°C
1
Meas.
current
500μA
250μA
500μA
250μA
500μA
250μA
500μA
250μA
250μA
250μA
250μA
250μA
500μA
250μA
500μA
250μA
500μA
250μA
500μA
250μA