Configuration Level 1 (C1-Level) - JUMO LOGOPRINT 500 Operating Manual

Printing recorder with text printing and led dot-matrix display
Table of Contents

Advertisement

4.2.3

Configuration level 1 (C1-level)

Plot status
Measurement input Configuration level 1
Parameter
Configuration level 1
Plot status
X Plot status
Status:
Measurement input
X Measurement input
Type:
Configuration level 1
Measurement input
Type: Resistance
thermometer
Characteristic:
Connection:
Unit:
(dimension)
Range start:
Range end:
Filter:
Configuration level 1
Measurement input
Type: thermocouple
Characteristic:
Unit:
(dimension)
1. The limits for range start and range end differ according to the characteristic
4 Configuration table
Value/selection
1 — 3 or 1 — 6
ON, OFF
1 — 3 or 1 — 6
RES.THERMOMETER
THERMOCOUPLE
POTENTIOMETER
RES. TRANSMITTER
VOLTAGE
CURRENT
SHUNT
Pt100, Pt100 JIS, Pt500,
Pt1000, Ni100, X1, X2,
2/3-wire
4-wire
°C, °F
-9999 to +9999
-9999 to +9999
Measurement span ≥ 15°C
0.0 — 50.0sec
L, J, U, T, K, E, N, S, R, B,
X1, X2
°C, °F
Description
Setting for each writing
channel: should the signal
trace be printed on the
chart and appear in the
statistical report or not.
Select the type of probe/
sensor and input signal,
determine the measure-
ment range + filter time
constant for each input
channel.
X1 = customer-specific
linearisation 1
X2 = customer-specific
linearisation 2
(with X1 or X2 : followed
immediately by the entry
for the filter parameter.
Determine the type of
connection for the
resistance thermometer.
Enter start of
1
measurement range
Enter end of
1
measurement range
Enter filter constant
X1 = customer-specific
linearisation 1
X2 = customer-specific
linearisation 2
(with X1 or X2 : followed
immediately by the entry
for the filter parameter)
57