AquaProbe 2

3 Electrical Installation

3.1 Connections

3.1.1 Sensor Terminal Box Connections (Remote Versions only)

Caution. (Remote versions)
Make connections only as shown.
Remove foil screens
Twist the three screen wires together and sleeve them.
Twist cable pairs together
Maintain Environmental Protection at all times.
Conduit connections must provide cable entry sealing.
Fig. 3.1 Sensor Terminal Box Connections (Remote Version)
Caution. With Belden Cable 8777, ensure that the black wires are not interchanged, and remain with
the associated twisted pair.
IM/AP
Issue 9
3 Electrical Installation
CABLE
ABB
Belden 8777
Violet
Blue
Green
Sleeved Ground
Yellow
Orange
Red
Green
Brown
Maximum Cable Lengths
≤ 250 m
≤ 80 m
White
Black
Red
Black
Black
17