ABB AX416 User Manual page 63

Single and dual input analyzers for ph/redox orp
Hide thumbs Also See for AX416:
Table of Contents

Advertisement

Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
-----
Cal. Output 1
16000
O1: Adjust 4mA
7200
O1: Adjust 20mA
Cal. Output 1
-----
Cal. Output 2
7200
O1: Adjust 20mA
Cal. Output 2
IM/AX4PH
Issue 9
Calibrate Output 1
Note. When adjusting the 4 and 20mA outputs, the display reading is unimportant and
is used only to indicate that the output is changing when the
pressed.
See below.
Cal. Output 2
Adjust 4mA
Use the
Note. The analog output range selected in
affect the reading.
Adjust 20mA
Use the
Note. The analog output range selected in
affect the reading.
See below.
Cal. Output 2
Monitoring pH
Redox (ORP)
Operating Page - see section 2.3, page 6.
Dual Redox
pH/Redox (ORP)
Calibrate Output 2
Note. Output 2 calibration is identical to Output 1 calibration.
Option board fitted and analog features enabled - continued on next page.
Cal. Output 3
Option board fitted, additional features disabled - continued on next page.
Option Board
Option board not fitted - continued on next page.
Alter Fact. Code
Option board fitted and analog features enabled - continued on next page.
Cal. Output 3
Option board fitted, additional features disabled - continued on next page.
Option Board
Option board not fitted - continued on next page.
Alter Fact. Code
Monitoring pH
Redox (ORP)
Operating Page - see section 2.3, page 6.
Dual Redox
pH/Redox (ORP)
and
keys to set the milliammeter reading to 4mA.
and
keys to set the milliammeter reading to 20mA.
7 CALIBRATION
and
(Section 5.6) does not
Configure Outputs
(Section 5.6) does not
Configure Outputs
keys are
61

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax436Ax460Ax466Ax468

Table of Contents