Cal. Sensor A
No
Yes
-----
A: Abort Cal.
60
Temperature Span (150R)
Set the temperature simulator to 150Ω
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (1k0)
Set the temperature simulator to 1000Ω
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (1k5)
Set the temperature simulator to 1500Ω
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (2k0)
Set the temperature simulator to 2000Ω
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (5k0)
Set the temperature simulator to 5000Ω
The display returns automatically to
recorded.
Abort Calibration
Select Yes or No
selected:
Yes
A:T.Zero (100R)
completion of
before
and continues.
CAL. SENSOR A
completion of
after
page.
selected - calibration continues from the point at which the
No
Cal. Sensor A
frame - calibration advances to
A:mV Span (+1V)
frame - the display returns to the
A:mV Span (+1V)
7 CALIBRATION
once a stable and valid value is
A:T.Zero (100R)
Calibrate Sensor A
key was pressed.